Quality Policy
Career Opportunities
Contact Us
Directions to ISS
Comments & Suggestions
Gregorio Weber Award
Internal Use
06/06/2007: ISS Awarded NIH Grant
03/03/2007: ISS Announces 2007 Gregorio Weber Award Winner
02/18/2006: ISS Announces 2006 Gregorio Weber Award Winner
09/01/2005: ISS Achieves ISO 9001:2000 Certification
01/05/2005: ISS Announces a New Version of OxiplexTS™
News Archive
Fluorescence Products
Alba FCS: Fluorescence Correlation SpectrometerAlba FLIM: Fluorescence Lifetime Imaging
PC1: Photon Counting Spectrofluorimeter
K2: Multifrequency Cross-Correlation Phase and Modulation Fluorometer
Chronos: Frequency-Domain Lifetime Spectrometer
ChronosBH: Time-Domain Lifetime Spectrometer
Modular Components for Building Spectrofluorometers
Phoenix: SLM Upgrade Packages
Software
Request Literature for Fluorescence Products
Biomedical Products
OxiplexTSImagent
Request Literature for Biomedical Products
Fluorescence Lifetime Standards
Lifetime Data of Selected Fluorophores
Excitation and Emission Wavelengths of Fluorophores
Fluorescent Probes Data Table
Fluorescence Standards for LEDs and Laser Diodes
Application & Technical Notes
Measurement Examples
Selected Publications
Laboratories & Societies
Anisotropy Decay Measurements
Ewald Terpetschnig
ISS, Inc.
Introduction
Knowledge of dynamic fluorescence parameters such as the fluorescence lifetime or the rotational correlation time is crucial for the quantitative interpretation of fluorescence data for quenching, energy transfer or polarization [1]. The lifetime is the average time a fluorophore spends in the excited state before returning to the ground state by emitting a photon. It is calculated as the time when the initial intensity has decayed by a factor of 1/e. Similarly the rotational correlation time is the time at which the initial anisotropy has decayed to 1/e of its original value.
Methods
As there are 2 methods to measure the lifetime there are also two ways to measure the rotational correlation times [2]:
-
Time-domain method (TD)
In time domain one measures the time-dependent intensity decays of the polarized components Iv and Ih with vertical excitation. These polarized decays are then used to calculate the time-dependent anisotropy decay:


The anisotropy decay r(t) is then analyzed to determine which model is most consistent with the data.
-
Frequency-domain method (FD)

In frequency domain the sample is excited with amplitude-modulated light and the emission is observed at parallel and perpendicular polarizer settings. There are two parameters, which characterize the anisotropy decay in FD:
Differential phase angle between the perpendicular and parallel components of the emission:

and the amplitude ratio between the parallel and perpendicular amplitudes of the modulation:

The amplitude ratios are often expressed as modulated anisotropy because of its higher information content:

rω is the frequency dependent anisotropy and it approaches ro at high modulation frequencies while at low modulation frequency it approaches the steady state anisotropy value. The differential phase angles Δω are approx. Lorentzian in shape on the log-frequency scale. Most fluorescent dyes show single correlation times in polar solvent systems even though they are not spherical molecules. This is due to hydrogen bonding between solvent and the fluorescent molecules.

Measurement Examples
We have measured several anisotropy decays on Chronos™ and found the measurement of time-resolved parameters to be fast and easy. In FD there is no need to measure the lifetime separately as the FD analysis provides information of both – the lifetimes and the rotational correlation times of the investigated system.


For other examples please go to: http://www.iss.com/products/k2/examples/index.html
In summary frequency domain measurements provide excellent resolution of complex anisotropy decays and this can be considered one of the strengths of the FD method [3].
One possible reason for this is that FD enables direct measurement of the difference between the polarized components, while in time-domain the polarized components are measured separately and then the anisotropy decay is calculated. The other explanation is that FD uses 2 parameters that are directly measured (Δω and Λω) and therefore has a better resolution potential [3].
References
- Jameson, D.M. and Sawyer, W.H. Fluorescence anisotropy applied to biomolecular interactions. Meth. Enzym. 246, 284-300 (1995).
- Lakowicz, J.R. (1999), Principles of Fluorescence Spectroscopy, 2nd Edition, Kluwer Academic/Plenum Publishers, New York.
- Lakowicz, J.R. (1999), Principles of Fluorescence Spectroscopy, 2nd Edition, Kluwer Academic/Plenum Publishers, New York, page 356, last paragraph.