Quality Policy
Career Opportunities
Contact Us
Directions to ISS
Comments & Suggestions
Gregorio Weber Award
Internal Use
06/06/2007: ISS Awarded NIH Grant
03/03/2007: ISS Announces 2007 Gregorio Weber Award Winner
02/18/2006: ISS Announces 2006 Gregorio Weber Award Winner
09/01/2005: ISS Achieves ISO 9001:2000 Certification
01/05/2005: ISS Announces a New Version of OxiplexTS™
News Archive
Fluorescence Products
Alba FCS: Fluorescence Correlation SpectrometerAlba FLIM: Fluorescence Lifetime Imaging
PC1: Photon Counting Spectrofluorimeter
K2: Multifrequency Cross-Correlation Phase and Modulation Fluorometer
Chronos: Frequency-Domain Lifetime Spectrometer
ChronosBH: Time-Domain Lifetime Spectrometer
Modular Components for Building Spectrofluorometers
Phoenix: SLM Upgrade Packages
Software
Request Literature for Fluorescence Products
Biomedical Products
OxiplexTSImagent
Request Literature for Biomedical Products
Fluorescence Lifetime Standards
Lifetime Data of Selected Fluorophores
Excitation and Emission Wavelengths of Fluorophores
Fluorescent Probes Data Table
Fluorescence Standards for LEDs and Laser Diodes
Application & Technical Notes
Measurement Examples
Selected Publications
Laboratories & Societies
Absorption Measurements on PC1™
Yevgen Povrozin and Ewald Terpetschnig
ISS, Inc.
Introduction
The ISS absorption accessory enables measuring of the wavelength-dependent optical density (OD) of a sample on PC1™ or K2™ spectrofluorimeters. The transmitted intensity follows the Lambert-Beer law for non-scattering, optically homogeneous non-aggregated samples [1]. Highly scattering samples, aggregates or fluorescent samples require additional correction. This accessory is installed in place of the sample compartment in K2™ or PC1™ spectrofluorimeters [2].
Figure 1. Absorption Accessory - Side View.
Instrumentation
The absorption spectra shown below were measured on PC1™, the photon-counting spectrofluorimeter from ISS. PC1™ features parallel beam geometry for reliable polarization measurements. Polarization measurements can be performed in the L or T-format. Vinci – Multidimensional Fluorescence Spectroscopy, a comprehensive and flexible fluorescence analysis software package, also enables instrument control and data acquisition directly from the PC.
Figure 2. Schematic drawing of PC1™, the photon-counting spectrofluorimeter from ISS.
Installing the Absorption Accessory
Although the standard use of the accessory will be for the left emission channel of the instrument (filter holder channel), the accessory can be used also with acquisition through the right emission channel (monochromator channel). In the following, we will assume that the accessory will be utilized with the left emission channel.
Please keep the left emission shutter closed. Turn the left emission PMT voltage temporarily off. This protects the detector from receiving an excessive amount of room light.
Remove the standard sample holder from the instrument. Remove the four 4/40 flat head screws holding the front of the sample compartment to the instrument. Also remove the two 8/32 socket head screws located next to the cooling can, under the emission lenses. Slide the turret halfway out and carefully disconnect all cable connectors. Remove the excitation channel lens holder and the two collecting lenses. Insert the absorption measurements accessory into the instrument.
Insert in the excitation channel the lens holder supplied with the accessory (focal length 200 mm). Move the excitation monochromator to 500 nm wavelength. Open the excitation shutter: the light beam is reflected via the first mirror onto the second and, from there, it will cross the cuvette holder. If not, adjust the mirrors in order to make the beam pass through the optical center of the cuvette holder (Figure 1 red).
Absorption Measurements
For this type of measurements it is necessary to use slits with a 1 mm bandwidth (8 nm) or less in the excitation monochromator and to insert neutral density filter with an OD=3 in the filter holder of the excitation channel. Finally the iris located in the excitation channel behind the sample compartment of the instrument needs to be closed to its full extent. All these operations are meant to reduce the light intensity reaching the sample and eventually the light detector. The level of PMT counts should be around 100K. It is important to keep the PMT gain knob position at its maximum (10).
The absorption measurement is performed in two steps:
- Measurement of the blank (Cuvette containing pure solvent) is collected setting the excitation monochromator to the selected wavelength;
- Measurement on the sample
Calculation of the OD:
The symbols are defined as follows:
- ε = Molar extinction coefficient [M-1cm-1]
- c = concentration [M/liter]
- x = optical path [cm]
- I0= incident intensity (blank light intensity)
- I = light intensity after passing through the sample
The absorption coefficient, K=εc, can be derived once the optical path is known. Standard cuvettes have an optical path of 1 cm. Other cuvettes can have a lower (or higher optical path), which can be entered in the equation to calculate the extinction coefficient:
Absorption spectroscopy utilizes quantities, which can be related to the optical density and the absorption coefficient. The Absorbance A is defined as:
Analysis
After the measurement of the sample and the blank, the absorption spectrum is obtained using Vinci Analysis software. For the calculation of the absorption values in Vinci the following steps are required:
Open the sample transmission file in Vinci Analysis and choose Spectral and then Determine Absorption Spectrum in the menu which brings up the Compute Absorption Spectrum dialog box, which asks to select the 2nd blank transmission spectrum. Select the spectrum by double-clicking on it and then click OK. This will produce the background corrected absorption spectrum.
Note: Both the sample and blank data files require smoothing with Vinci Analysis before the actual absorption spectrum is calculated: Go to Math in Vinci Analysis, then choose Smooth and then Savitzky-Golay (or any other) for smoothing of the raw spectra.
Measurement Examples
In the following measurements we compared the ODs of Cy5-solutions that were obtained on PerkinElmer’s Lamda25 to those measured on PC1™ using the absorption accessory. The plots below show the spectra of these solutions measured on these two instruments as well as the calculated difference of their absorption maxima [in %].
Experimental conditions: excitation slits: 0.5 mm, lamp current: 18A, PMT gain: 10, OD=3 neutral density filter in excitation channel. The original data were smoothed by averaging with Savitzky-Golay by a factor of 6.
References
- H. Langhals, G. Abbt-Braun, F. H. Frimmel. Association of Humic Substances: Verification of Lambert-Beer Law. Acta hydrochimica et hydrobiologica 28 (6), 329 – 332 (2001).
- H. E. Rose. Breakdown of the Lambert-Beer Law. Nature 169, 287 - 288 (1952).