Quality Policy
Career Opportunities
Contact Us
Directions to ISS
Comments & Suggestions
Gregorio Weber Award
Internal Use
02/20/2010: ISS Announces 2010 Gregorio Weber Award Winner
02/28/2009: ISS Announces 2009 Gregorio Weber Award Winner
02/02/2008: ISS Announces 2008 Gregorio Weber Award Winner
06/06/2007: ISS Awarded NIH Grant
03/03/2007: ISS Announces 2007 Gregorio Weber Award Winner
News Archive
Fluorescence Products
Alba FCS: Fluorescence Correlation SpectrometerAlba FLIM: Fluorescence Lifetime Imaging
PC1: Photon Counting Spectrofluorimeter
K2: Multifrequency Cross-Correlation Phase and Modulation Fluorometer
ChronosFD: Frequency-Domain Lifetime Spectrometer
ChronosBH: Time-Domain Lifetime Spectrometer
Modular Components
Phoenix: SLM Upgrade Packages
Accessories
Software
Request Literature for Fluorescence Products
Biomedical Products
OxiplexTSImagent
Request Literature for Biomedical Products
Fluorescence Lifetime Standards
Lifetime Data of Selected Fluorophores
Excitation and Emission Wavelengths of Fluorophores
Fluorescent Probes Data Table
Fluorescent Proteins Table
Fluorescence Standards for LEDs and Laser Diodes
Application & Technical Notes
Measurement Examples
Selected Publications
Laboratories & Societies
Fluorescence Lifetime (FLT)
Ewald Terpetschnig1 and David M. Jameson2
1ISS Inc.
2Department of Cell and Molecular Biology
John A. Burns School of Medicine
1960 East-West Rd.
University of Hawaii, HI 96822-2319
Principles
The fluorescence lifetime is a measure of the time a fluorophore spends in the excited state before returning to the ground state by emitting a photon [1]. The lifetimes of fluorophores can range from picoseconds to hundreds of nanoseconds. A list of the some commonly used fluorophores and their fluorescence lifetimes are given in Table 1.
| Fluorophore | Lifetime [ns] | Excitation Max [nm] | Emission Max [nm] | Solvent |
| ATTO 655 | 3.6 | 655 | 690 | Water |
| Acridine Orange | 2.0 | 500 | 530 | PB pH 7.8 |
| Alexa Fluor 488 | 4.1 | 494 | 519 | PB pH 7.4 |
| Alexa Fluor 647 | 1.0 | 651 | 672 | Water |
| BODIPY FL | 5.7 | 502 | 510 | Methanol |
| Coumarin 6 | 2.5 | 460 | 505 | Ethanol |
| CY3B | 2.8 | 558 | 572 | PBS |
| CY3 | 0.3 | 548 | 562 | PBS |
| CY5 | 1.0 | 646 | 664 | PBS |
| Fluorescein | 4.0 | 495 | 517 | PB pH 7.5 |
| Oregon Green 488 | 4.1 | 493 | 520 | PB pH 9 |
| Ru(bpy)2(dcpby)[PF6]2 | 375 | 458 | 650 | Water |
| Pyrene | > 100 | 341 | 376 | Water |
| Indocyanine Green | 0.52 | 780 | 820 | Water |
| Rhodamine B | 1.68 | 562 | 583 | PB 7.8 |
Table 1. Commonly used fluorophores and their fluorescence lifetimes.


It = α e-t/τ
Where It is the intensity at time t, α is a normalization term (the pre-exponential factor) and τ is the lifetime.
Knowledge of the excited state lifetime of a fluorophore is crucial for quantitative interpretations of numerous fluorescence measurements such as quenching, polarization and FRET.
Excited state lifetimes have traditionally been measured using the “time domain” method or the “frequency domain” method.
-
Time-domain method
In the time domain method, the sample is illuminated with a short pulse of light and the intensity of the emission versus time is recorded. Originally, these short light pulses were generated using flashlamps that had widths on the order of several nanoseconds.
Modern laser sources can now routinely generate pulses with widths on the order of picoseconds or shorter.

If the decay is a single exponential and the lifetime is long compared to the exciting light, then the lifetime can be determined directly from the slope of the curve. If the lifetime and the excitation pulse width are comparable, some type of deconvolution method must be used to extract the lifetime.
Great effort has been expended on developing mathematical methods to “deconvolve” the effect of the exciting pulse shape on the observed fluorescence decay (see, for example, many chapters in [2]). With the advent of very fast laser pulses these deconvolution procedures became less important for most lifetime measurements, although they are still required whenever the lifetime is of comparable duration to the light pulse.
-
Frequency-domain method
tan φ = ωτ





Thus using the phase shift and relative modulation one can determine a phase lifetime τp and a modulation lifetime τM.
τP (ω1) < τP (ω2) if ω1 > ω2
The differences between τp and τM and their frequency dependence form the basis of the methods used to analyze for lifetime heterogeneity, i.e., the component lifetimes and amplitudes.

Analysis

In addition to decay analysis using discrete exponential decay models, one may also choose to fit the data to distribution models. In this case, it is assumed that the excited state decay characteristics of the emitting species actually results in a large number of lifetime components. Shown below is a typical lifetime distribution plot for the case of a single tryptophan containing protein – Human Serum Albumin.

The plots show the frequency response curves (phase and modulation vs. modulation frequency) for Human Serum Albumin (left). The excitation source was a 300-nm UV-LED; the emission was collected through a WG320 high-pass filter at a temperature of 20°C. Lifetime analysis was performed using a Lorentzian distribution (center at 5.4 ns, width = 2.9 ns, fractional distribution = 98%) and a second discrete component (t = 0.51 ns and fractional contribution = 0.02%). For a review of HSA lifetime studies see [3].
The distribution shown here is Lorentzian, but depending on the decay kinetics of the system, different types of distributions, e.g., Gaussian, or asymmetric distributions (Planck), may be utilized. This approach to lifetime analysis is described in [4].
Applications
Fluorescence Lifetime Assays:
The fluorescence lifetime (FLT) has been widely utilized for the characterization of fluorescence species and in biophysical studies of proteins, e.g. the distances between particular amino-acid residues by Foerster Resonance Energy-Transfer (FRET). FLT is a parameter that is mostly unaffected by inner filter effects, static quenching and variations in the fluorophore concentration. For this reason FLT can be considered as one of the most robust fluorescence parameters, and therefore it is advantageous in clinical and high throughput screening (HTS) applications where it is necessary to discriminate against the high background fluorescence from biological samples. Also FLT offers more leverage with regards to multiplexing. The ability to discriminate between two fluorophores with similar spectra but different lifetimes is another way to increase the number of parameters to be measured (see, for example [5]).
Several mechanisms can be utilized for the development of lifetime-based assays. There are the simple binding assays, where binding of 2 components (one being fluorescently labeled) is accompanied by a FLT-change. Another scenario would be a quench-release type assay where the quenched species has low but finite fluorescence but is initially present in large excess. If the fluorescence compound is released (binding to a complementary DNA strand (Molecular Beacon) or by an enzymatic reaction) the lifetime of the system increases. Finally, FLT is a powerful method to measure energy transfer efficiency in FRET (fluorescence resonance energy transfer) assays, circumventing the issue of spectral cross talk between donor and acceptor, by using a non-fluorescent acceptor.
Fluorescence Lifetime Sensing:
Most of the fluorescence sensors and assays that are in use today are based on intensity measurements. Though these methods are easier to implement they lack robustness and they require frequent calibration [6]. Many difficulties that are associated with intensity-based measurements can be circumvented using lifetime-based measurements. Lifetime-based measurements have the advantage that they are independent of the fluorescence intensity. In past 10 years many probes that exhibit analyte-sensitive fluorescence lifetime changes have been identified and characterized. Some of these probes are listed in Table 2. For a detailed discussion on lifetime-based sensing we refer you to the book chapter “Lifetime-based Sensing” in [6].
Fluorescence Lifetime Imaging:
Fluorescence lifetimes also offer opportunities in fluorescence microscopy where the local probe concentration cannot be controlled. FLIM allows image contrast to be created based on the fluorescence lifetime of a probe at each point of the image. Typical examples are the mapping of cell parameters such as pH, ion concentrations or oxygen saturation by fluorescence quenching, fluorescence resonance energy transfer (FRET), or photon-induced energy transfer (PET). Examples of biological applications of lifetime imaging technology include scanning of tissue surfaces, photodynamic therapy, DNA chip analysis, skin imaging and others (see, for example [7]).
| Fluorescent Probes | Mean Lifetime [ns] | Absorption Max [nm] | Emission Max [nm] | |||
| free | bound | free | bound | free | bound | |
| a) Calcium Probes | ||||||
| Fura-2 | 1.09 | 1.68 | 362 | 335 | 500 | 503 |
| Indo-1 | 1.4 | 1.66 | 349 | 331 | 482 | 398 |
| Ca-Green | 0.92 | 3.66 | 506 | 506 | 534 | 534 |
| Ca-Orange | 1.20 | 2.31 | 555 | 555 | 576 | 576 |
| Ca-Crimson | 2.55 | 4.11 | 588 | 588 | 610 | 612 |
| Quin-2 | 1.35 | 11.6 | 356 | 336 | 500 | 503 |
| b) Magnesium Probes | ||||||
| Mg-Quin-2 | 0.84 | 8.16 | 353 | 337 | 487 | 493 |
| Mg-Green | 1.21 | 3.63 | 506 | 506 | 532 | 532 |
| c) Potassium Probe | ||||||
| PBFI | 0.52 | 0.59 | 350 | 344 | 546 | 504 |
| d) Sodium Probe | ||||||
| Sodium Green | 1.13 | 2.39 | 507 | 507 | 532 | 532 |
| e) pH Probes | ||||||
| SNAFL-1 | 1.19 | 3.74 | 539 | 510 | 616 | 542 |
| Carboxy-SNAFL-1 | 1.11 | 3.67 | 540 | 508 | 623 | 543 |
| Carboxy-SNAFL-2 | 0.94 | 4.60 | 547 | 514 | 623 | 545 |
| Carboxy-SNARF-1 | 1.51 | 0.52 | 576 | 549 | 638 | 585 |
| Carboxy-SNARF-2 | 1.55 | 0.33 | 579 | 552 | 633 | 583 |
| Carboxy-SNARF-6 | 1.03 | 4.51 | 557 | 524 | 635 | 559 |
| Carboxy-SNARF-X | 2.59 | 1.79 | 575 | 570 | 630 | 600 |
| Resorufin | 2.92 | 0.45 | 571 | 484 | 587 | 578 |
| BCECF | 4.49 | 3.17 | 503 | 484 | 528 | 514 |
Table 2. Spectral properties (absorption and emission maxima) and mean lifetimes of common ion-probes.
Books and Book Chapters related to Fluorescence Lifetime:
- 1. Lakowicz, J.R. (1999). Principles of Fluorescence Spectroscopy, 2nd Edition, Kluwer Academic/Plenum Publishers, New York.
- 2. Valeur, B. (2002). Molecular Fluorescence. Wiley-VCH Publishers.
- 3. Herman B. (1998). Fluorescence Microscopy, 2nd Edition, Springer-Verlag, New York.
- 4. Baeyens W.R.G., de Keukeleire, D., Korkidis, K. (1991). Luminescence techniques in chemical and biochemical analysis, M. Dekker, New York.
- 5. Jameson, D. M. and Hazlett, T.L. (1991). Time-Resolved Fluorescence in Biology and Biochemistry, in Biophysical and Biochemical Aspects of Fluorescence Spectroscopy (Dewey, Ed.) Plenum Press, New York.
References:
- Weber, G. in Hercules, D.M. Fluorescence and Phosphorescence Analysis. Principles and Applications, Interscience Publishers (J. Wiley & Sons), New York, pp. 217-240 (1966).
- Cundall, R.B. and Dale, R.E. (Eds.). Time-Resolved Fluorescence Spectroscopy in Biochemistry and Biology (Nato Advanced Science Institutes Series. Series a, Life Sciences; Vol. 69, Plenum Pub Corp, New York (1983).
- Helms, M.K., Petersen, C.E., Bhagavan, N.V., Jameson, D.M., Time-resolved fluorescence studies on side-directed mutants of human serum albumin. FEBS letters, 408, 67-70 (1997).
- Alcala, J. R., Gratton E. and Prendergast, F.G., Fluorescence lifetime distributions in proteins. Biophys. J. 51, 597-604 (1987).
- Gratton E. and Jameson, D.M., New approach to phase and modulation resolved spectra. Anal. Chem. 57, 1694-1697 (1985).
- Szmacinski H. and Lakowicz, J.R., Topics in Fluorescence Spectroscopy: Vol. 4. Probe Design and Chemical Sensing Lakowicz, J.R. (Ed.), Plenum Press, New York, (1994).
- Clegg, R. M. Holub, O., and Gohlke, C., Fluorescence lifetime-resolved imaging: measuring lifetimes in an image. Methods Enzymol. 360, 509-542 (2003).