Alba FLIM™ Specifications

Instrument Features

  • Dual channel acquisition
  • Separate pinholes for each channel for higher resolution
  • Computer-controlled alignment of the confocal pinhole and optics
  • Choice between scanning mirrors or piezo-controlled XYZ stage
  • Single- and multi-photon excitation
  • Powered by Vista, a user-friendly software package for the acquisition of FLIM, FRET, RICS, and FCS
  • Fully upgradable to Alba FCS™

Measurements

  • Fluorescence Lifetime Images (FLIM)
  • Fluorescence Resonance Energy Transfer (FRET) efficiency images
  • Fluorescence images
  • Raster imaging correlation spectroscopy (RICS)
  • Steady-state polarization and ratio imaging
  • Time courses (kinetics)

Image Parameters Acquired by Alba FLIM™:

  • Image area and resolution
  • XY and XZ sections
  • Speed
  • Step size = 50 nm
  • Pixels numbers: user selectable from 20 to 1024

Optical Unit

  • Light Sources:
    • Single photon lasers housed in a laser launcher with computer-controlled beam expander, laser intensity and shutter; or,
    • Multi-photon excitation with computer-controlled beam expander, laser intensity and shutter
  • Optics:
    • Nikon TE2000 inverted microscope with lens revolver for objectives, bottom and side ports and computer-controlled XY-stage for sample slides and for 384-well plates.
    • Available for Leica, Olympus and Zeiss microscopes
    • Objectives:
      • Air objectives with 20X, 40X, 60X magnification and 1.5-8.1 working distances
      • Oil immersion objectives, 1.4 NA and 60X (standard); other aperture available
      • Water immersion objectives,1.2 NA 60X (standard), with coverslip correction (for 0.15-0.18 coverslip); other apertures available
    • Dichroic Filters:
      • For single-photon excitation: 1-, 2-, 3-band filters
      • For multi-photon excitation
      • Polarizer:
      • Cube beam splitter, wavelength range: 450-1100 nm; extinction ratio: 10,000:1 at +/- 3degrees
  • Imaging Stage:
    • XYZ piezo-controlled stage, 100x100x50 µm with 5 nm step resolution.
    • Optional Coarse Microscope Stage:
      • Large distance movement (100x100x10 mm)
      • Manual stage or
      • Stepper motor controlled XYZ stage
      • Both stages accommodate
        • Microwell plates
        • Petri dishes
        • Coverslip
  • Light Detectors:
    • Photomultiplier tubes
      • Spectral range: 250 nm - 850 nm
      • Dark counts: less than 200 cp
    • Optional APD detectors for RICS
  • Laser Scanner Unit:
    • Galvo-controlled mirrors

Modulation Unit

  • The modulation and detection unit includes computer-controlled frequency synthesizers (1 -300 MHz) and an RF amplifier.

Data Acquisition

  • Data acquisition board
    • Proprietary, 32 bit PCI bus, two-channel card.
  • Sampling rate
    • 200 KHz
  • Input/Output triggers
    • For monitoring temperature, pH, laser light source fluctuation and starting measurement.

Software: Vista-FCS and Confocal Imaging Microscopy

  • Operating system
    • Windows 2000/XP
  • Image format
    • Export to ImageJ, MetaMorph
    • Plots export to png, gif, jpeg, bitmap formats

Instrument

  • Computer
    • Intel-type CPU, Windows XP operating system
  • Power Requirements
    • Universal power input: 110-240 V, 50/60 Hz, 400 VAC
  • Dimensions
    • 885 mm (L) x 600 mm (W) x 330 mm (H)
  • Weight
    • 40 kg

Return to Alba FLIM™